Biotek ISA601PRO XL International Electrical Safety Analyzer
Price: Call
Product Description
- PREOWNED: Very Good Condition
- SALES LEAD TIME: 14 Days (Unit will be sent out for fresh calibration certificate after purchase)
- WARRANTY: 12 Months
- ACCESSORIES: Power Cord, Test Leads, Operator Manual, Calibration Certificate, 2 Rolls of paper
FEATURES
- Multiple enclosure-leakage points
- Multiple patient-applied-part types
- Power ON/OFF delay
- DC only for patient- and auxiliary-leakage tests
- User-programmable test sequences
- Manual, auto, step, and computer-control modes
- ASCII data transfer
- Memory for up to 1000 device-information records
- EN60601-1, EN601010-1, and AAMI test loads (user selectable)
SPECIFICATIONS:
Voltage: Range: 0 to 300 V True RMS
Accuracy: DC-100 Hz+/- 1.5% of reading +/- LSD
Insulation Resistance: Range: 0.5 to 400.0 Mohms
Accuracy: +/- 5% of reading +/- LSD
Current Consumption: Range: 0 to 15 A ac True RMS
Accuracy: +/- 5% of reading +/- LSD
Mains on Applied Part: Applied Voltage: > or = 110 % of mains voltage
Accuracy: +/- 2% of reading +/- 6 micro A
Supply Voltage: 90 to 265 VAC, Autoswitching
Protective Earth Resistance: Range: 0.000 to 2.999 ohms
Accuracy: +/- 5% of reading +/- 4 mohms (1A, 10A, and 25A test currents)
IEC601-1 and AAMI Range: 0 – 8000 microA True RMS
Leakage Currents Accuracy: (per IEC601-1 or AAMI filter)
-DC – 1 kHz +/- 1% of reading +/- 1 microA
-1 to 100kHz +/- 2% of reading +/- 1 microA
-100 kHz to 1 MHz +/- 5% of reading +/- 1 microA
DC-Only Frequency Response: DC – 5 Hz (approx)
IEC1010-1 Leakage Current: Range: 0 – 16000 microA True RMS
(For some leakage tests, displayed range may be 0 – 9999 when using IEC1010 test load)
Accuracy: (per IEC1010-1 figure A.1 filter)
-DC – 1 kHz +/- 2% of reading +/- 2 microA
-1 to 100 kHz +/- 4% of reading +/- 2 microA
-100 kHz to 1 MHz +/- 10% of reading +/- 10 microA
DC-Only Frequency Response: DC – 5 Hz (Approx)
ECG Simulation and ECG Complex: 30, 60, 120, 180, 240 BPM
Performance Testing: Performance Pulse: 30, 60, BPM, 63 ms pulse width (600 – 700 micros rise and fall time)
Sine Waves: 10, 40, 50, 60, 100 Hz
Square Waves: 0.125, 2.000 Hz (50% duty cycle)
Triange Wave: 2 Hz, 2 mV
Arrhythmia: Atrial Fibrillation, Atrial Flutter, Atrial Tachycardia, Idioventricular, PVC1, R on T, Run
Ventricular Fibrillation, Ventricular Tachycardia
Performance: +/- 2% of reading for rate or +/- 5% of reading for amplitude, fixed at 1 mV peak
on a Lead II ECG connection (except for triange wave, which is 2mV peak to peak)
Dimensions: 16.62 in L x 11.75 in W x 5.56 in H
42.2 cm L x 30 cm W x 14.1 cm H
Weight: 17lb
7.7kg