Netech EXPMT 2000 External Pacemaker Tester
Brand: Netech Corp
Product Code: Netech EXPMT 2000 External Pacemaker Tester
Availability: Indent 1-2 bulan
Harga :CALL
Product Description
OVERVIEW:
EXPMT 2000, Pacemaker Analyzer is the ultimate Pacer tester designed for evaluating and testing of all Atrial / Ventricular Pacemakers including Transthoracic Pacers.
The New EXPMT 2000, External Pacemaker Analyzer, has evolved from the best selling EXPMT 100 Pacer Analyzer. It will test and verify all the functions of any external pacemaker including Transvenous, Transthoracic, and A-V Sequential models.
EXPMT 2000, Pacemaker Analyzer with its new look, new features, new controls, and new display combine to make it the easiest to use and the most comprehensive pacer analyzer available.
The operation of this Pacemaker Analyzer is menu driven via eight tactile keys and a large four line LCD display. Test measurements are simultaneously presented on the display. Five ECG snaps are standard along with an RS 232 serial port that allows for printing botha heading and test measurements on a separate printer. Information may also be sent to a Palm Pilot. The External Test Load has twelve user selectable loads from 100 to 1000 Ohms for use with transcutaneous pacers.
- FEATURES:
- Easy to Use, Fast & Accurate
- 12 Selectable Test Loads
- 5 ECG Test Leads
- RS 232 Interface and a dedicated Print Key for simplicity
- Accurate Test results, Pulse Rate, Width, Current (amplitude)
- Demand Sensitivity Tests, Sensed and Paced RefactoryTests
- A-V Interval, Pulse Energy
- 60 Hz/ 50 Hz Immunity Tests
- SPECIFICATIONS:
- Transthoracic Pacer
Test Load: 50 Ohm fixed12 Selectable step from 100 to 1000 ohms with EXPMT Load SelectorCurrent: 0.1 to 25 mA
- Accuracy: +/-10%, 1 to 10 mA,
+/-5%, 10 to 200 mA- Pulse Rate: 30 to 800 PPM
- Accuracy: +/-1% or 2 ppm
- Pulse Width: 0.5 to 80 ms
- Accuracy: +/-10%, 0.5 to 5 ms,
+/-5%, 5 to 25 ms,
+/-1%, 25 to 80 ms- Sensed and Paced Refractory Period: 25 to 500 ms
- Accuracy: +/-10 ms
- 50/60 Hz Immunity Test: 0 to 10.6 mV in 0.7 mV steps
- Transvenous Pacer
- Test Load: 500 Ohms +/- 1%
- Current: 0.1 to 25 mA
- Accuracy: +/-10%, 0.1 to 10 mA
+/-5%, 1 to 25 mA- Pulse Rate: 30 to 800 PPM
- Accuracy: +/-1% or 2 ppm
- Pulse Width: 0.5 to 80 ms
- Accuracy: +/-10%, 0.5 to 5 ms
+/-5%, 5 to 25 ms
+/-1%, 25 to 80 ms- Sensed and Paced Refractory
- Period: 25 to 500 ms
- Accuracy: +/-10 ms
- 50/60 Hz Immunity Test: 0 to 102 mV in 6.8 mV steps
- Demand Sensitivity
- Waveforms: SQR, TRI, and SSQ
- Amplitude: 0 to 100 mV.
- Width: 10, 25, 40, 100, and 200 ms
- Pacer Rate: 30 to 120 ppm
- A – V Interval: 0 to 999 ms
- General
- Display: 4 Line LCD Display
- Housing: ABS Plastic
- Power Requirements: 9V Battery or AC Operation
- Physical Dimensions: 7.75 x 4.0 x 1.5 inches
(19.7 x 10.2 x 3.8 cm)- Weight: 1 lbs (0.9 kg)
- Environmental:
Operating Temperature: 59degrees to 95degrees F (15degrees to 35degrees C)